MinION Mk1C (MC-112691) Final report

04 Nov 22, 15:14 — 05 Nov 22, 15:15 · st2022 · St3-4-20221104 · MC-112691
Protocol run ID: c813f0e6-4ca4-4a41-96a3-27686190393b

DATA OUTPUT

Estimated bases
254.68 Mb
Data produced
4.67 GB
Reads generated
355.94 k
Estimated N50
848

BASECALLING

Reads called
100%
Bases called (min Q score: 9)
141.49 Mb
97.17 Mb

RUN DURATION

Elapsed time

24 hours of 24 hours

Run status

Finished

Flow Cell type
FLO-FLG001
Flow Cell ID
AHR740
Kit type
SQK-LSK110
Specified run length
24hrs
Active channel selection
On
Pore scan freq.
1.5hrs
Bias voltage (initial)
-20mV
Bias voltage (final)
-50mV
Reserved pores
Off
Basecalling
High-accuracy model
FAST5 output
vbz_compress
FAST5 reads per file
10000
FASTQ output
compress
FASTQ reads per file
10000
BAM output
Off
Bulk file output
Off
Data location
/data/./st2022/St3-4-20221104/20221104_1514_MC-112691_AHR740_c813f0e6
MinKNOW
22.05.8
Bream
7.1.3
Configuration
5.1.5
Guppy
6.1.5
MinKNOW Core
5.1.0

The read length graph shows the total number of bases vs the read length. The longest 1% of strands are classified as outliers, and excluded to allow focus on the main body of data.

Legend

Basecalled
Estimated
Estimated N50
848
% Basecalled
100%

The longest 1% of strands are classified as outliers, and aggregated into groups to show their relative amounts.

Read length (kb) Aggregated reads (Mb)
3.75 - 4.75 1.05
4.75 - 5.75 0.21
5.75 - 6.75 0.1
6.75 - 7.75 0.03
7.75 - 8.375 0.03

The cumulative output shows the total amount of bases or reads sequenced over time by your device.

Bases

Legend

Estimated
Predicted total number of bases, prior to basecalling
Passed
Bases equal to or above the quality score threshold.
Failed
Bases below the quality score threshold.

Reads

Legend

Total
Total number of reads, including passed, failed and skipped.
Passed
Reads equal to or above the quality score threshold.
Failed
Reads below the quality score threshold.
Skipped
Reads that will not be basecalled. Post run basecalling is possible.

The quality score is calculated as basecalling is performed on your device. Quality score is measured on the logarithmic PHRED scale. Reads that fall below the minimum value of 9 will be classified as failed reads. You can alter the accepted minimum quality score in MinKNOW.

Legend

Mode
The most frequent quality score of reads in the run.
Spread
The spread of quality scores, found by calculating full width half maximum.
Min. quality score
Minimum quality score to be accepted as a passed read.

Quality score low

This can be due to the translocation speed being out of the accepted range, check the translocation speed graph on this report. If the translocation speed is in the accepted range, please see the user guide for more info.

The Pore activity graph shows the performance of your sample as it is being sequenced during a run.

Legend

Sequencing
Pore currently sequencing
Pore
Pore available for sequencing
Recovering
Inactive pore that may recover
Inactive
Pore unavailable for sequencing
Unclassified
Pore not assigned a status

General

Some commonly seen issues are excess pores classified as Recovering, Open Pore, or Free Adapter. To find out what advice is applicable for your run, visit the user guide.

A Pore scan is performed at configurable time intervals to determine the current status of pores within channels on a Flow Cell. For this run a Pore scan is performed every 1.5hrs.

Legend

Active wells
Wells available for sequencing
Inactive wells
Wells unavailable for sequencing

High proportion Unavailable

Possible contaminants in library blocking the pore. Consider using the Flow Cell Wash Kit, and reloading a library. 

High proportion Inactive

If localised to one area of the Flow Cell, this could indicate that an air bubble has been introduced during the flushing/loading steps. If inactivity is spread across the Flow Cell this could be caused by improper loading of the library, please refer to the user guide for further support.

The translocation speed is the rate at which DNA/RNA travels through pores as it is being sequenced.

Legend

Median
75% quartile
25% quartile
Accepted range

Low speed

Check that the Flow Cell is within the target temperature range.

The temperature of the Flow Cell over the run time.

Legend

Measured
Target

Out of range

Check that the Flow Cell is correctly seated and firmly pushed down into the device. Ensure ambient temperature is always within the specified range for your device in the user guide.

SYSTEM MESSAGES

System messages are a record of the events that occurred in the time covered by this report.

None
None
Disk space · 2022-11-04T15:14:54.941977Z
Disk /data has 653 GB space remaining
Waiting for temperature · 2022-11-04T15:15:03.176480Z
Waiting up to 300 seconds for temperature to stabilise at 35.0°C
Starting · 2022-11-04T15:15:30.629363Z
Starting sequencing procedure
Message · 2022-11-04T15:15:30.695320Z
Performing Channel Scan
Message · 2022-11-04T15:19:03.164669Z
Channel scan for flow cell AHR740 has found a total of 43 pores. 43 pores available for immediate sequencing
Message · 2022-11-04T16:50:01.768028Z
Performing Channel Scan
Message · 2022-11-04T16:53:56.640891Z
Channel scan for flow cell AHR740 has found a total of 35 pores. 35 pores available for immediate sequencing
Message · 2022-11-04T18:25:05.805619Z
Performing Channel Scan
Message · 2022-11-04T18:29:00.887761Z
Channel scan for flow cell AHR740 has found a total of 29 pores. 29 pores available for immediate sequencing
Message · 2022-11-04T19:59:51.323147Z
Performing Channel Scan
Message · 2022-11-04T20:03:46.504842Z
Channel scan for flow cell AHR740 has found a total of 19 pores. 19 pores available for immediate sequencing
Message · 2022-11-04T21:34:37.604884Z
Performing Channel Scan
Message · 2022-11-04T21:38:32.843777Z
Channel scan for flow cell AHR740 has found a total of 15 pores. 15 pores available for immediate sequencing
Message · 2022-11-04T23:09:23.405193Z
Performing Channel Scan
Message · 2022-11-04T23:13:18.522873Z
Channel scan for flow cell AHR740 has found a total of 13 pores. 13 pores available for immediate sequencing
Message · 2022-11-05T00:44:21.318990Z
Performing Channel Scan
Message · 2022-11-05T00:48:16.539Z
Channel scan for flow cell AHR740 has found a total of 10 pores. 10 pores available for immediate sequencing
Message · 2022-11-05T02:19:05.119253Z
Performing Channel Scan
Message · 2022-11-05T02:23:00.356175Z
Channel scan for flow cell AHR740 has found a total of 9 pores. 9 pores available for immediate sequencing
Message · 2022-11-05T03:53:47.699252Z
Performing Channel Scan
Message · 2022-11-05T03:57:42.898766Z
Channel scan for flow cell AHR740 has found a total of 5 pores. 5 pores available for immediate sequencing
Message · 2022-11-05T05:28:30.184949Z
Performing Channel Scan
Message · 2022-11-05T05:32:25.347136Z
Channel scan for flow cell AHR740 has found a total of 7 pores. 7 pores available for immediate sequencing
Message · 2022-11-05T07:03:11.236180Z
Performing Channel Scan
Message · 2022-11-05T07:07:06.424Z
Channel scan for flow cell AHR740 has found a total of 3 pores. 3 pores available for immediate sequencing
Message · 2022-11-05T08:37:51.856222Z
Performing Channel Scan
Message · 2022-11-05T08:41:47.126778Z
Channel scan for flow cell AHR740 has found a total of 4 pores. 4 pores available for immediate sequencing
Message · 2022-11-05T10:12:35.210229Z
Performing Channel Scan
Message · 2022-11-05T10:16:30.382402Z
Channel scan for flow cell AHR740 has found a total of 3 pores. 3 pores available for immediate sequencing
Message · 2022-11-05T11:47:15.435070Z
Performing Channel Scan
Message · 2022-11-05T11:51:10.721305Z
Channel scan for flow cell AHR740 has found a total of 2 pores. 2 pores available for immediate sequencing
Message · 2022-11-05T13:21:56.443725Z
Performing Channel Scan
Message · 2022-11-05T13:25:51.832768Z
Channel scan for flow cell AHR740 has found a total of 3 pores. 3 pores available for immediate sequencing
Message · 2022-11-05T14:56:37.431945Z
Performing Channel Scan
Message · 2022-11-05T15:00:33.002170Z
Channel scan for flow cell AHR740 has found a total of 2 pores. 2 pores available for immediate sequencing
Message · 2022-11-05T15:15:45.224930Z
The sequencing run has finished, but basecalling may continue