MinION Mk1C (MC-112691) Final report

06 Nov 22, 06:18 — 07 Nov 22, 06:19 · 2022st · st1-2 · MC-112691
Protocol run ID: 2b3b06be-1ee4-4c6b-a362-a718c662fca9

DATA OUTPUT

Estimated bases
228.73 Mb
Data produced
3.29 GB
Reads generated
179.48 k
Estimated N50
1.54 kb

BASECALLING

Reads called
100%
Bases called (min Q score: 9)
100.25 Mb
106.49 Mb

RUN DURATION

Elapsed time

24 hours of 24 hours

Run status

Finished

Flow Cell type
FLO-FLG001
Flow Cell ID
AHW051
Kit type
SQK-LSK110
Specified run length
24hrs
Active channel selection
On
Pore scan freq.
1.5hrs
Bias voltage (initial)
-20mV
Bias voltage (final)
-20mV
Reserved pores
Off
Basecalling
High-accuracy model
FAST5 output
vbz_compress
FAST5 reads per file
10000
FASTQ output
compress
FASTQ reads per file
10000
BAM output
Off
Bulk file output
Off
Data location
/data/./2022st/st1-2/20221106_0618_MC-112691_AHW051_2b3b06be
MinKNOW
22.05.8
Bream
7.1.3
Configuration
5.1.5
Guppy
6.1.5
MinKNOW Core
5.1.0

The read length graph shows the total number of bases vs the read length. The longest 1% of strands are classified as outliers, and excluded to allow focus on the main body of data.

Legend

Basecalled
Estimated
Estimated N50
1.54 kb
% Basecalled
100%

The longest 1% of strands are classified as outliers, and aggregated into groups to show their relative amounts.

Read length (kb) Aggregated reads (kb)
3.25 - 11.25 972.17
11.25 - 19.25
None
19.25 - 27.25
None
27.25 - 29.75 30.28

The cumulative output shows the total amount of bases or reads sequenced over time by your device.

Bases

Legend

Estimated
Predicted total number of bases, prior to basecalling
Passed
Bases equal to or above the quality score threshold.
Failed
Bases below the quality score threshold.

Reads

Legend

Total
Total number of reads, including passed, failed and skipped.
Passed
Reads equal to or above the quality score threshold.
Failed
Reads below the quality score threshold.
Skipped
Reads that will not be basecalled. Post run basecalling is possible.

The quality score is calculated as basecalling is performed on your device. Quality score is measured on the logarithmic PHRED scale. Reads that fall below the minimum value of 9 will be classified as failed reads. You can alter the accepted minimum quality score in MinKNOW.

Legend

Mode
The most frequent quality score of reads in the run.
Spread
The spread of quality scores, found by calculating full width half maximum.
Min. quality score
Minimum quality score to be accepted as a passed read.

Quality score low

This can be due to the translocation speed being out of the accepted range, check the translocation speed graph on this report. If the translocation speed is in the accepted range, please see the user guide for more info.

The Pore activity graph shows the performance of your sample as it is being sequenced during a run.

Legend

Sequencing
Pore currently sequencing
Pore
Pore available for sequencing
Recovering
Inactive pore that may recover
Inactive
Pore unavailable for sequencing
Unclassified
Pore not assigned a status

General

Some commonly seen issues are excess pores classified as Recovering, Open Pore, or Free Adapter. To find out what advice is applicable for your run, visit the user guide.

A Pore scan is performed at configurable time intervals to determine the current status of pores within channels on a Flow Cell. For this run a Pore scan is performed every 1.5hrs.

Legend

Active wells
Wells available for sequencing
Inactive wells
Wells unavailable for sequencing

High proportion Unavailable

Possible contaminants in library blocking the pore. Consider using the Flow Cell Wash Kit, and reloading a library. 

High proportion Inactive

If localised to one area of the Flow Cell, this could indicate that an air bubble has been introduced during the flushing/loading steps. If inactivity is spread across the Flow Cell this could be caused by improper loading of the library, please refer to the user guide for further support.

The translocation speed is the rate at which DNA/RNA travels through pores as it is being sequenced.

Legend

Median
75% quartile
25% quartile
Accepted range

Low speed

Check that the Flow Cell is within the target temperature range.

The temperature of the Flow Cell over the run time.

Legend

Measured
Target

Out of range

Check that the Flow Cell is correctly seated and firmly pushed down into the device. Ensure ambient temperature is always within the specified range for your device in the user guide.

SYSTEM MESSAGES

System messages are a record of the events that occurred in the time covered by this report.

None
None
Disk space · 2022-11-06T06:18:36.041284Z
Disk /data has 648 GB space remaining
Waiting for temperature · 2022-11-06T06:18:43.796327Z
Waiting up to 300 seconds for temperature to stabilise at 35.0°C
Starting · 2022-11-06T06:19:11.557350Z
Starting sequencing procedure
Message · 2022-11-06T06:19:11.608831Z
Performing Channel Scan
Message · 2022-11-06T06:22:43.686188Z
Channel scan for flow cell AHW051 has found a total of 34 pores. 34 pores available for immediate sequencing
Message · 2022-11-06T07:53:42.065207Z
Performing Channel Scan
Message · 2022-11-06T07:57:38.149320Z
Channel scan for flow cell AHW051 has found a total of 25 pores. 25 pores available for immediate sequencing
Message · 2022-11-06T09:28:35.041165Z
Performing Channel Scan
Message · 2022-11-06T09:32:33.410862Z
Channel scan for flow cell AHW051 has found a total of 23 pores. 23 pores available for immediate sequencing
Message · 2022-11-06T11:03:31.677806Z
Performing Channel Scan
Message · 2022-11-06T11:07:30.777917Z
Channel scan for flow cell AHW051 has found a total of 22 pores. 22 pores available for immediate sequencing
Message · 2022-11-06T12:38:23.033271Z
Performing Channel Scan
Message · 2022-11-06T12:42:22.401065Z
Channel scan for flow cell AHW051 has found a total of 21 pores. 21 pores available for immediate sequencing
Message · 2022-11-06T14:13:19.283937Z
Performing Channel Scan
Message · 2022-11-06T14:17:18.593240Z
Channel scan for flow cell AHW051 has found a total of 15 pores. 15 pores available for immediate sequencing
Message · 2022-11-06T15:48:07.565189Z
Performing Channel Scan
Message · 2022-11-06T15:52:02.598192Z
Channel scan for flow cell AHW051 has found a total of 10 pores. 10 pores available for immediate sequencing
Message · 2022-11-06T17:22:49.464728Z
Performing Channel Scan
Message · 2022-11-06T17:26:44.960242Z
Channel scan for flow cell AHW051 has found a total of 10 pores. 10 pores available for immediate sequencing
Message · 2022-11-06T18:57:33.244204Z
Performing Channel Scan
Message · 2022-11-06T19:01:28.612522Z
Channel scan for flow cell AHW051 has found a total of 7 pores. 7 pores available for immediate sequencing
Message · 2022-11-06T20:32:20.155174Z
Performing Channel Scan
Message · 2022-11-06T20:36:15.351777Z
Channel scan for flow cell AHW051 has found a total of 6 pores. 6 pores available for immediate sequencing
Message · 2022-11-06T22:07:05.955194Z
Performing Channel Scan
Message · 2022-11-06T22:11:01.404239Z
Channel scan for flow cell AHW051 has found a total of 7 pores. 7 pores available for immediate sequencing
Message · 2022-11-06T23:41:48.201303Z
Performing Channel Scan
Message · 2022-11-06T23:45:44.017816Z
Channel scan for flow cell AHW051 has found a total of 6 pores. 6 pores available for immediate sequencing
Message · 2022-11-07T01:16:29.311848Z
Performing Channel Scan
Message · 2022-11-07T01:20:24.626313Z
Channel scan for flow cell AHW051 has found a total of 2 pores. 2 pores available for immediate sequencing
Message · 2022-11-07T02:51:09.853434Z
Performing Channel Scan
Message · 2022-11-07T02:55:05.027167Z
Channel scan for flow cell AHW051 has found a total of 3 pores. 3 pores available for immediate sequencing
Message · 2022-11-07T04:25:52.583953Z
Performing Channel Scan
Message · 2022-11-07T04:29:48.196358Z
Channel scan for flow cell AHW051 has found a total of 3 pores. 3 pores available for immediate sequencing
Message · 2022-11-07T06:00:33.401129Z
Performing Channel Scan
Message · 2022-11-07T06:04:28.899436Z
Channel scan for flow cell AHW051 has found a total of 3 pores. 3 pores available for immediate sequencing
Message · 2022-11-07T06:19:26.040019Z
The sequencing run has finished, but basecalling may continue